Oven & Furnace

Calibration of Oven & Furnace.
Laboratory calibration capability range covered :

  • 0 ~ 1000 °C

Method 1 – Reference sensor comparison with oven sensor
For the above method, one reference sensor is placed within close proximity to the oven sensor.
Recorded maximum, minimum & mean temperature of reference are reported.

Method 2 – Reference sensor placed in one desire location of the oven.
For the above method, one reference sensor is placed in one location(normally Center/middle) of the oven as instructed by customer.
Recorded maximum, minimum & mean temperature of reference are reported.

Method 3 – 9 reference sensors are placed throughout the volume of the oven.
For the above method, 9 reference sensor are placed in the 8 corners & center of the oven
Characteristics of chamber such as temperature fluctuation, temperature gradient, variation in space, maximum, minimum & mean temperature are reported.

Note:
User to inform us their oven set temperature, measurement duration & method required.
If other measurement requirement is required, do let us know.