Chiller & Freezer

Calibration of Freezer & Chiller,
(applicable to walk-in Freezer & Chiller)
Laboratory calibration capability range covered :

  • -35 °C ~ 15 °C

As different industry customer require different kind of measurement method for their freezer & chiller. Do let us know your requirement. We offer several way measurement.

Method 1 – Reference sensor comparison with freezer/chiller sensor.
For the above method, one reference sensor is placed within close proximity to the freezer/chiller.
Recorded maximum, minimum & mean temperature of reference are reported.

Method 2 – Reference sensor placed in one desire location of the freezer/chiller.
For the above method, one reference sensor is placed in one location(normally Center/middle) of the freezer/chiller as instructed by customer.
Recorded maximum, minimum & mean temperature of reference are reported.

Method 3 – 9 reference sensors are placed throughout the volume of the freezer/chiller.
For the above method, 9 reference sensor are placed in the 8 corners & center of the freezer/chiller.
Characteristics of chamber such as temperature fluctuation, temperature gradient, variation in space, maximum, minimum & mean temperature are reported.

Note:
User to inform us their freezer/chiller set temperature, measurement duration & method required.
If other measurement requirement is required, do let us know.